      • 產品名稱:Vermason 222547校準器

      • 產品型號:用于校準手腕帶和工鞋接地測試儀
      • 產品廠商:其它品牌
      • 產品文檔:
      你添加了1件商品 查看購物車
      簡單介紹:
      Vermason 222547校準器 Calibration Unit for Vermason Wrist Strap and Footwear Tester用于校準手腕帶和工鞋接地測試儀適配儀器:222504, 222506,222507, 222562, 222563, 222564, 222565,Vermason(DESCO子品牌)
      詳情介紹:

      Vermason 222547校準器 Calibration Unit for Vermason Wrist Strap and Footwear Tester
      用于校準手腕帶和工鞋接地測試儀
      適配儀器:222504, 222506,222507, 222562, 222563, 222564, 222565
      Vermason(DESCO子品牌)

      Operation
      The Tester Calibration Unit was specifically designedfor use in calibration of Vermason brand equipment.The following procedures will not cover adjustment of
      the test equipment. For detailed information regardingadjustment of specific Vermason brand testers, pleasecontact Desco Europe Customer Service.

      Frequency of recalibration should be based on thecritical nature of those ESD sensitive items handled andthe risk of failure for the ESD protective equipment andmaterials. In general, we recommend that calibration be
      performed annually.

      Testing the Compact Tester
      WRIST STRAP CIRCUIT
      1. Connect the 10 mm lead from the Calibration Unitto the wrist strap stud on the Compact Tester.
      2. Select 715K FAIL LOW on the Calibration Unit.
      3. Press and hold the Calibration Unit’s test probeon the Compact Tester’s test plate. The testershould indicate a FAIL LOW condition.
      4. Select 785K PASS LOW on the Calibration Unitand repeat the test. The tester should indicate aPASS condition.
      5. Select 33M PASS HIGH on the Calibration Unitand repeat the test. The tester should indicate aPASS condition.
      6. Select 37M FAIL HIGH on the Calibration Unitand repeat the test. The tester should indicate aFAIL HIGH condition.

      FOOTWEAR CIRCUIT
      1. Connect the 10 mm lead from the Calibration Unitto the 10 mm footwear stud on the Compact Tester.
      2. Select 715K FAIL LOW on the Calibration Unit.
      3. Press and hold the Calibration Unit’s test probeon the Compact Tester’s test plate. The testershould indicate a FAIL LOW condition.
      4. Select 785K PASS LOW on the Calibration Unitand repeat the test. The tester should indicate aPASS condition.
      5. Select 33M PASS HIGH on the Calibration Unitand repeat the test. The tester should indicate aPASS condition.
      6. Select 37M FAIL HIGH on the Calibration Unit andrepeat the test. The tester should indicate aFAIL HIGH condition

      姓名:
      電話:
      您的需求:
       
      久久精品国产亚洲精品| 亚洲午夜久久久久久尤物| 亚洲福利在线观看| 亚洲日韩精品无码专区加勒比☆| 亚洲高清无码专区视频| 亚洲国产精品无码AAA片| 亚洲伦理中文字幕| 亚洲人成电影网站国产精品 | 亚洲精品天堂无码中文字幕| 亚洲看片无码在线视频| 亚洲精品一区二区三区四区乱码 | 亚洲kkk4444在线观看| 亚洲最大AV网站在线观看| 国产成人精品日本亚洲直接| 国产精品久久久亚洲| 色噜噜亚洲精品中文字幕| 亚洲人成人网站在线观看| 最新亚洲人成网站在线观看| 色噜噜噜噜亚洲第一| 精品久久久久亚洲| 国产亚洲精品第一综合| 国产99久久亚洲综合精品| 国产青草亚洲香蕉精品久久| 亚洲专区中文字幕| 激情综合亚洲色婷婷五月APP| 亚洲精品一品区二品区三品区| 亚洲A∨精品一区二区三区下载| 亚洲精品第一国产综合精品| 亚洲卡一卡2卡三卡4卡无卡三| 亚洲福利视频导航| 亚洲福利电影一区二区?| 亚洲男同帅GAY片在线观看| 色偷偷亚洲第一综合网| 国产综合成人亚洲区| 亚洲福利精品电影在线观看| 亚洲色偷偷综合亚洲AV伊人蜜桃| 亚洲色偷偷综合亚洲AV伊人蜜桃 | 亚洲日韩精品无码一区二区三区| 久久亚洲高清综合| 亚洲国产成人高清在线观看| 亚洲天堂免费在线视频|